Whether it’s monitoring glaucoma risk, refining treatment plans, or providing reassurance during routine checks, measuring intraocular pressure (IOP) is an integral part of daily practice. Yet how we capture this vital data is evolving.
We’re all familiar with the gold standard – Goldmann applanation tonometry. As the most established form of contact tonometry, it’s long been the go-to, with little else coming close in terms of accuracy.
However, newer rebound devices, such as iCare’s IC100 and IC200 tonometers, are reshaping expectations around comfort, portability and access. For today’s clinicians, there is now a reliable alternative to Goldmann, one backed up by a wide variety of studies.
Does it completely replace Goldmann? No, not at all. Instead, rebound tonometers offer a convenient alternative that is portable, accurate and, most importantly, more comfortable for patients.
Accuracy and agreement with Goldmann
The strength of Goldmann lies in its repeatability when used correctly. Numerous studies confirm it remains the “gold standard”, which is why it underpins referral thresholds. But head-to-head studies with modern rebound devices show that iCare has matured into a robust clinical alternative in many scenarios.
For example, a 2016 study comparing the IC100, its predecessor TA01i, and GAT in patients with and without glaucoma found that IC100 typically read slightly lower than Goldmann, with mean differences of around −4 mmHg and limits of agreement between −10 and +2 mmHg. Importantly, differences were influenced by central corneal thickness and curvature, highlighting the need to interpret results in the context of pachymetry and corneal biomechanics.
Follow-up work on the IC100 has reinforced its reliability. A study of healthy and post-LASIK eyes found excellent repeatability (intraclass correlation coefficients >0.9) and good agreement with GAT (ICC ~0.8), suggesting that even in surgically altered corneas, the device performs consistently. Other comparative studies show broadly similar patterns: good overall correlation, but wider variability at very high or very low pressures.
Data confirms that, while Goldmann remains the definitive tool, modern rebound tonometers provide readings close enough for confident use in most clinical contexts, provided clinicians remain aware of corneal and IOP-range effects.
Patient comfort and clinical workflow
For patients, the distinction between applanation and rebound is often stark. Goldmann requires anaesthetic and fluorescein (which nobody enjoys). By contrast, the iCare devices require no drops and involve only a virtually undetectable tap of the probe.
In practical terms, this difference improves tolerance in children, anxious patients and those seen at home. Clinical studies with the IC200 in paediatric cohorts have shown reliable readings with far greater acceptability than applanation, making IOP assessment feasible in children who would otherwise be untestable.
Patients are often less reluctant and quicker to process when using iCare tonometers, making it easier to manage workflow efficiency in busy practices.
Clinical applications
The choice between Goldmann and rebound tonometry is not black and white.
In hospital glaucoma clinics, Goldmann (or Perkins) continues to be essential for diagnosis, progression analysis and medico-legal purposes. In community optometry, domiciliary care and paediatrics, however, IC100 and IC200 offer clear advantages: they are fast, well-tolerated, and portable.
The IC200 in particular opens up additional clinical applications. Its ability to measure patients accurately in both upright and supine positions has been shown to reveal posture-related IOP changes, useful in research and in complex glaucoma cases where diurnal or positional variation may drive disease progression.
Devices such as the iCare Home2, which share the same rebound principle, also provide opportunities for home monitoring, allowing clinicians to capture diurnal IOP profiles that Goldmann alone cannot provide.
A side-by-side comparison
As a quick overview:
- iC100 is ideal for seated or standing patients
- iC200 is also used for supine patients in wards or theatres
- iCare Home2 is offered for self-monitoring between clinic visits
And, to help you compare the two, we’ve created a quick reference:
| Feature | Goldmann Applanation Tonometry (GAT) | iCare Rebound Tonometry |
| Accuracy | Gold standard when used correctly | Correlates closely with GAT but is not a full replacement |
| Equipment | Slit lamp, prism, fluorescein, anaesthetic | Handheld, portable device |
| Patient comfort | Requires anaesthetic, can sting | No anaesthetic, generally well tolerated |
| Hygiene | Reusable prism; must be disinfected | Single-use probes – no sterilisation required |
| Training | Considerable operator skill required | Minimal training needed |
| Use case | Clinic-based | Flexible – clinics, wards, domiciliary visits |
| Limitations | Not mobile; impacted by corneal thickness/curvature | Less accepted for legal thresholds and not accepted for some hospital referrals. Also, accuracy may vary with abnormal corneas |
What’s right for your clinic?
Goldmann applanation tonometry remains the gold standard, but it no longer needs to stand alone.
The latest generation of rebound tonometers show how innovation can complement tradition rather than replace it. iCare devices widen access to reliable IOP data, supporting early detection and long-term monitoring in settings where conventional applanation may be impractical.
For clinicians balancing patient comfort and accuracy, combining GAT with rebound technology offers the best of both worlds.
If you’d like guidance on how iCare systems can enhance IOP assessment in your clinic, we’re here to help.
Contact us at [email protected] or call 0121 458 6800 to discuss what iCare rebound tonometry can offer to your clinic.
Featured products
-
iCare EIDON – Ultra-Widefield Confocal System
Enquire for pricing- Increased field of view up to 200°
- Multiple imaging modalities: TrueColor, blue, red, RedFree, infrared
- Capable of imaging through cataracts and media opacities
- Dilationfree operation through pupils as small as 25 mm
-
iCare DRSplus – TrueColor Confocal Fundus Imaging System
Enquire for pricing- Multiple imaging modalities: redfree, infrared, external, stereo views
- No dilation needed, ensuring comfortable patient experience
- Fast, fully automated imaging through pupils as small 2.5 mm
- Remote Exam feature allows exams to be performed remotely
-
Enquire for pricing
- High-resolution Scheimpflug imaging captures 3D anterior segment
- Contact-free measurement offers rapid, reproducible results
- Detailed corneal and chamber maps enhance diagnostics
- Fast Screening Report simplifies clinical decision layout
-
Enquire for pricing
- 5‑in‑1 system integrates OCT, fundus, biometry, topography
- 80,000 A‑scans/sec delivers rapid, stable 3D imaging
- Single‑shot true‑colour fundus captures clear retinal anatomy
- Detailed OCT‑A visualises microvascular structures without dye
-
Optomed Lumo – Portable Fundus Camera
Enquire for pricing- Non‑mydriatic retinal imaging, undilated pupils
- Large 5″ intuitive touchscreen display
- Wireless DICOM image transfer connectivity
- High‑sensitivity 12 MP sensor optics
-
TowardPi – BMizar 400KHz Full-Range SS-OCT/A
Enquire for pricing- 400kHz swept-source engine: 400,000 A-scans per second
- Ultra-widefield OCTA: 30 × 25mm single-capture, no stitching
- 10 billion voxels of ultra-high-resolution imaging data
- AI-powered quantification across retina, choroid, and anterior segment
-
Enquire for pricing
- Combines tomography, biometry, refraction, wavefront, retroillumination scanning
- Contact-free axial length measurement for precise biometric data
- Whole-eye wavefront analysis reveals optical aberration profiles
- Supports comprehensive diagnostics for surgery, screening, management
-
Enquire for pricing
- Combined ultra-high-resolution OCT and Scheimpflug imaging
- Single contact-free scan captures whole corneal microstructure
- Aligned data enhances diagnostic accuracy and reproducibility
- Supports advanced corneal assessment and surgical planning
-
iCare EIDON – Ultra-Widefield Confocal System
Enquire for pricing- Increased field of view up to 200°
- Multiple imaging modalities: TrueColor, blue, red, RedFree, infrared
- Capable of imaging through cataracts and media opacities
- Dilationfree operation through pupils as small as 25 mm
-
iCare DRSplus – TrueColor Confocal Fundus Imaging System
Enquire for pricing- Multiple imaging modalities: redfree, infrared, external, stereo views
- No dilation needed, ensuring comfortable patient experience
- Fast, fully automated imaging through pupils as small 2.5 mm
- Remote Exam feature allows exams to be performed remotely
-
£2,995.00 + VAT
- No need for anesthetic drops, air, or calibration
- Automated Measurements Sequence
- Quick IOP readings enhance clinical efficiency
- Simple operation with minimal training required
-
£4,495.00 + VAT
- 200° position freedom
- No need for anesthetic drops, air, or calibration
- Quick Measure for easier and faster IOP measurements
- Enhanced probe control improves handling and measurement precision
-
Enquire for pricing
- Patients measure IOP throughout daily activities independently easily
- Measurements also possible while in supine body position
- Cloud uploads share real world IOP data securely
- Doctors access accurate data supporting better treatment decisions
-
£90.00 + VAT
- Box of 100 probes
- Ensures minimal contact and maximum patient comfort
- Individually packaged for hygienic, contaminationfree measurements
- Simple insertion and removal for efficient workflow
-
iCare ST500 – Slit Lamp Mounted Rebound Tonometer
Enquire for pricing- Slit lamp integration fits most lamps for seamless measurement
- Fast, accurate, and repeatable IOP measurements
- Stable patient positioning for easier measurement
- No need for anesthetic drops, air, or calibration
-
Enquire for pricing
- Highresolution corneal topography and keratometry
- Noninvasive tear film and dry eye analysis
- Meibography for gland visualisation and assessment
- Contact lens fitting and keratoconus support
-
Enquire for pricing
- Ultracompact slitlamp camera system
- Highresolution photo and video capture
- Swivelling prism for easy operation
- Integrated software with measurements and annotations
-
Tomey MR-6000 Multifunction Unit
Enquire for pricing- All‑in‑one ophthalmic diagnostics for six eye examinations
- Automatic alignment with rapid mode switching workflow
- Integrated dry eye and corneal topography analysis
- Tonometry corrected by pachymetry for accurate IOP
-
£94.50 + VAT
- Precise eyelid retraction and eversion for accurate assessment
- Supports meibography, dry eye, and lid cleaning procedures
- Disposable medicalgrade construction ensures hygiene and patient safety
- Integrated ruler and shaped edges assist clinical examinations efficiently
-
Tomey MR-6000 Multifunction Unit
Enquire for pricing- All‑in‑one ophthalmic diagnostics for six eye examinations
- Automatic alignment with rapid mode switching workflow
- Integrated dry eye and corneal topography analysis
- Tonometry corrected by pachymetry for accurate IOP
-
Enquire for pricing
- Compact refraction workstation optimized for efficient eye exams
- Smooth height adjustment supports precise patient alignment
- Integrated storage and cable management for organized setup
- Durable design ideal for busy optometry clinics
-
Huvitz HS-7000/HS-7500 40x Mag Slit Lamp
Enquire for pricing- Five‑step magnification drum provides versatile visual detail
- Galilean optics enhance clarity for accurate clinical diagnosis
- Multiple built‑in filters improve contrast for specific tests
- Joystick with trigger simplifies positioning and image capture
-
Huvitz HDR-9000 Digital Refractor
Enquire for pricing- 21‑Point refraction method ensures systematic exam precision
- Wireless Wi‑Fi communication connects to clinical devices
- Dual cross‑cylinder lenses support accurate astigmatic testing
- Tiltable body and height control enhance patient comfort
-
Enquire for pricing
- Measures refraction, axial length, keratometry in one
- Software guides complete myopia assessment and reporting process
- Quick, non-contact measurement quality supports painless exams
- Reliable results help detect myopia progression early
-
Enquire for pricing
- Combines tomography, biometry, refraction, wavefront, retroillumination scanning
- Contact-free axial length measurement for precise biometric data
- Whole-eye wavefront analysis reveals optical aberration profiles
- Supports comprehensive diagnostics for surgery, screening, management